Troubleshooting

X
The device does not contain any parts which may be repaired by the user!
Motor not
Check
running
• Is there mains/battery power present?
• Is the device switched on?
Insufficient
Check
vacuum
• Check general condition of the device according to VIII, 1
• The setting of the vacuum regulator
• The condition of the tubings
• Tight seal of valve disks inside safetyChamber
• Membrane is clean and has no tears
• If canister, lid and seals have tears, brittle areas or blockages
Running
Calibrating the rechargeable battery => Chapter VII
time from
battery
insufficient
Too loud
Check position and seal of valve disks inside SafetyChamber
If the fault cannot be corrected, contact your service center

Replacing the rechargeable battery

Xl
Disconnect the pump from the mains supply
1) open lid
2) Replace the rechargeable battery (use only Medela battery)
3) Plug in battery
4) Close lid
Corresponding illustrations are to be found on the back flap
Routine check
Xll
The Clario Home Care Pump requires no maintenance, but please perform the functional tests
before each use. There are no parts in the Clario motor-unit that can be repaired by the user.