Frontier™ 5000 Series Multi
5 MAINTENANCE
5.1 Maintenance and cleaning
5.1.1
General Care
Maintenance of the centrifuge is confined to keeping the rotor, the rotor chamber, and the rotor accessories clean
as well as to regularly lubricating the rotor insert bolts of a swing out rotor (if available).
Suitable lubricant can be order under this order number: 30314586
Own lubricants containing molycote and graphite are not allowed.
Please pay special attention to anodized aluminum parts (if available). Breakage of rotors can be caused even by
slight damage.
In case of rotors, buckets or tube racks getting in touch with corrosive substances the concerned spots have to be
cleaned carefully.
Corrosive substances are for instance: alkalis, alkaline soap solutions, alkaline amines, concentrated acids,
solutions containing heavy metals, water-free chlorinated solvents, and saline solutions. e.g. salt water, phenol,
halogenated hydrocarbons.
Cleaning – centrifuges, rotors, accessories
5.1.2
•
Turn the device off and disconnect it from the power supply before you begin any cleaning or disinfecting.
Do pour liquids into the housing interior.
•
Do not spray disinfectant on the device.
•
Thorough cleaning not only has its purpose in hygiene but also in avoiding corrosion due to pollution.
•
In order to avoid damaging anodized parts such as rotors, reduction plates etc., only pH-neutral detergents
with a pH-value of 6-8 may be used for cleaning. Alkaline cleaning agents (pH-value > 8) must not be used.
After cleaning, please ensure all parts are dried thoroughly, either by hand or in a hot-air cabinet (max.
temperature + 50°C).
•
It is necessary to coat anodized aluminum parts with anti-corrosion oil regularly in order to increase their
life- spans and reduce corrosion predisposition.
•
Due to humidity or not hermetically sealed samples, condensate may be formed. The condensate has to be
removed from the rotor chamber with a soft cloth regularly.
NOTE!
The maintenance procedure must be repeated every 10 to 15 runs, or at least once a week.
•
Connect the unit to the power supply, after the equipment is completely dry.
•
Do not carry out disinfection with UV-, beta- and gamma-rays or other high energy radiation.
5.1.3
Centrifuge cleaning and disinfection
•
Open the lid before you turn off the unit. Disconnect it from the power supply.
•
Remove the rotor screw or nut by turning counterclockwise.
•
Remove the rotor.
•
For cleaning and disinfection of the unit and the rotor chamber use the above-mentioned cleaner.
•
Clean all accessible areas of the device and its accessories, including the power cord with a damp cloth.
•
Wash the rubber seals and rotor chamber thoroughly with water.
•
Rub the dry rubber seals with glycerol or talc to prevent these to becoming brittle. Other components of the
unit, e.g. motor shaft and rotor cone must not be greased.
•
Dry the motor shaft with a soft, dry, and lint-free cloth.
•
Control the unit and accessories for damage.
EN-26