EN-25
4.1.5
Call up operating data
In the mode "Basic Adjustments" you can call up the operating data of the centrifuge. Please proceed as
described under point 4.1.1 to enter this program mode. Press the key "Accel/Decel" (4). In the display "Acc/Dec"
(A-2) the word "Service" (M12) flashes.
With the touch wheel (1) the different information can be accessed:
A = previous starts of the centrifuge
H = previous operating hours
h = running time of the motor
S = software version
E = list of previous error messages
F = Function of the Imbalance Sensor
U = Intermediate Circuit Voltage in Volt
P = Operation of keyboard
d = Hardware Version
The list of the last 99 error messages can be looked over by pressing the key "RPM | RCF" (3) and scroll through it
by the touch wheel (1). The respective error codes appear in the display "RPM | RCF" (A-1). Please refer to "Table
4: error messages" (see APPENDIX).
To change back to normal program mode again, switch off the centrifuge for a short period.
A-1
3
4
M12
A-2
Figure 33
Frontier™ 5000 Series Multi
1