TROUBLESHOOTING

8
Fault
The scale does not
display a weight al-
though a load was
placed on it
Scale switches itself
on, e.g. during trans-
port
The scale displays a
weight after transport
or a battery replace-
ment although no
load has been placed
on it
does not appear
0.0
before weighing
displayed before
----
weighing
50 • seca 878
CAUTION!
Personal injury
If you find faults or deviations during the function
check which you are unable to remedy with the aid
of the "Troubleshooting" section in this document,
you must not use the device.
► Have the device repaired by seca Service or an
authorized service partner.
► Follow the section entitled "Servicing" in this
document.
If faults occur when operating the device, first attempt to
remedy them yourself using the following table. If the fault
persists, contact seca Service or an authorized service
partner.
Cause
Scale is switched off or
without batteries
key pressed
start
Scale has determined an
incorrect zero point
Scale had load on it before
being switched on
Scale had load on it before
being switched on
Remedy
• Check that the scale is
switched on
• Check that batteries are
inserted
Set the On/off switch to
"O" when transporting
Switch off the scale, then
switch it back on again
• Remove the load from
the scale
• Switch off the scale, then
switch it back on again
• Remove the load from
the scale
• Switch off the scale, then
switch it back on again