Herunterladen Diese Seite drucken

Gaymar Plexus Medical Auto SureFloat Bedienungsanleitung Seite 7

Werbung

Auto SureFloat™
__________________________________________________________________________
8.0 Operator Control Panel
1.
ON/STANDBY switches Control Unit between On
and STANDBY modes.
2.
POWER INDICATOR LED indicates AC power is
present to the Control Unit.
3.
MODE SELECT Press repeatedly to toggle through
operating modes, Auto/Float, Auto/A/P,
Manual/Float and Manual/A/P.
3.1 AUTO MODE (Sensor cord from M1000S
mattress must be connected to control unit)
Mattress will automatically inflate to optimal level
based on information received from active sensors
in mattress. The indicator adjacent to the icon, will
light while automatic mode is active.
Note: Connecting the mattress sensor cord while
the unit is not in automatic mode will cause the
Control Unit to enter automatic mode immediately.
If Sensor cord from mattress becomes disconnected
from control unit while in Auto mode, alarm will
sound.
3.2 MANUAL MODE Mattress inflation level user-
determined by Comfort Adjust ( - ,+ ) setting. The
indicator adjacent to the icon will light while manual
mode is active.
3.3 FLOAT MODE all cells within mattress will be
uniformly inflated when this icon is illuminated.
3.4 AP MODE. (Alternating Pressure) Adjacent
cells within mattress will be alternately inflated and
deflated in this mode. The indicator adjacent to the
icon will light while AP mode is active. Note: AP
mode is temporarily suspended in max inflate mode.
4.
SOFT (-) decreases the inflation level of the bed
while Control Unit is in manual mode only. Pressing
once will decrease the setting by one comfort level.
Pressing and holding will rapidly change the setting.
5.
COMFORT CONTROL INDICATOR displays the
current inflation level setpoint.
6.
FIRM (+) increases inflation level of the bed while in
manual mode only. Pressing once will increase the
setting by one comfort level setting (~2.5mmHg).
Pressing and holding will rapidly change the setting.
7.
MAX INFLATE selects/deselects the max inflate
mode (maximum mattress firmness). The indicator
adjacent to the button will flash while max inflate
mode is active. After 30 minutes, max inflate mode
will cancel.
8.
LOCKOUT activates or cancels lockout mode.
Press and hold for three seconds to activate or
cancel mode. The indicator adjacent to the button
will light while lockout is active. In lockout mode,
only the MAX INFLATE button is operational. All
other buttons are temporarily inactive.
9.
ALARM SILENCE silences indefinitely, the audible
alarm.
10. POWER FAIL INDICATOR flashes when power
supplied from line cord is interrupted while the
Control Unit is operating. To avoid an inadvertent
power fail alarm, place the unit in STANDBY mode
before disconnecting power from Control Unit.
6
Operator Control Panel

Werbung

loading

Diese Anleitung auch für:

C1000msM1000s