Auto SureFloat™
__________________________________________________________________________
Automatic Mode - AP
1. If alternating low pressure therapy is also desired, press the Mode Select button until the indicators
adjacent to the Auto and A/P icons are illuminated.
2. In Auto – A/P mode, alternate cells within the mattress are inflated and deflated every 2½ minutes. At
the end of each 2½ minute cycle the system will evaluate the floatation level, (determined by the
automatic sensors within the mattress) and make comfort level adjustments as necessary before
resuming the A/P operation.
11.0 Transferring Patients In and Out of Bed
1. For patient transfer press the MAX INFLATE button. This will maximally inflate the mattress to provide a
firm and stable support surface (and suspend AP and automatic modes). Pressing the MAX INFLATE
button again will cancel this mode and return the mattress to previously chosen settings.
2. While in the max inflate mode, the indicator adjacent to the MAX INFLATE button will flash to remind the
caregiver that the system is in this state. After approximately 30 minutes, the Control Unit will
automatically cancel the max inflate mode and return to the previously chosen settings.
12.0 CPR (Cardio Pulmonary Resuscitation)
To deflate mattress for CPR:
1. Disconnect the dual quick-connect couplings of the mattress hose from the Control Unit. Mattress will
deflate in approximately 15-20 seconds.
2. Proceed with CPR procedures.
......
Instructions for Use / Transferring Patients / CPR
.
9