Care / Storage / Troubleshooting
__________________________________________________________________________
13.2 Storage
Control Unit
1. Check the AC power cord and plug for abrasions or excessive wear.
2. Coil power cord up and place both cord and Control Unit in plastic bag for storage.
Mattress
1. Check the air manifold of the mattress for kinks or breaks. Replace if necessary.
2. To deflate, disconnect hose assembly from Control Unit.
3. Coil the hose assembly loosely and roll the deflated air mattress starting at the head end. Use the
straps for containment. Place in plastic bag for storage.
14.0 Troubleshooting
Symptom:
Mattress does not inflate or is too soft.
Power loss indicator flashing and
audible alarm sounding
Indicator adjacent to ALARM
SILENCE button is flashing and
audible alarm is sounding.
Corrective Action
Make certain the air hoses are not kinked, cracked, or split.
Verify the hose couplings are fully engaged to controller. Verify
that all 20 air cells are connected to the air manifold within
mattress.
Power has been disconnected from unit while in an operational
mode. Restore power, put unit in STANDBY mode, then remove
power. Unit must always be returned to STANDBY mode before
disconnecting power.
If in AUTO mode, check the connection of the M1000S mattress
sensor cord to the jack on the side of the unit. Reconnect if
necessary.
Verify hose couplings are fully engaged to controller. Verify that
all 20 air cells are connected to the air manifold within mattress.
11
Auto SureFloat™