Herunterladen Inhalt Inhalt Diese Seite drucken

Operation Of The Minifg; Switch On The Device; How Do I Carry Out The Stimulation - Pierenkemper Mini FG Gebrauchsanweisung

Frequenzgenerator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6
3

Operation of the MiniFG

Switch on the Device

Switch on the device by means of the
Figure 1
key. The MiniFG will start with Program 1 (P1).

How do I carry out the stimulation?

Selection of the Program
You change to the next standard or user program by operating
the modification keys
programs 1 to 6 indicated with 'P' will run through initially. Then,
the user programs identified with 'U' will appear, as far as they exist.
When the final program has been reached, the device will spring
back to program 1 (P1) at the next keystroke.
The relevant operating parameters for each program will be shown
in the display:
Sinus
Rectangle
Positive – Negative
Positive
C
Continuing
P
Program (7-20-7-20-7min.)
F
6 Frequencies at 10min.
1
1.
Display
2.
Menu keys
3.
Modification keys
The intensity and the
different parameters can
2
be changed during
stimulation and in
editing mode by
means of these keys.
4.
On / Off – Button
5.
Battery compartment
6.
Output sockets
4
5
(Figure 1). During this, the standard
English
19

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis