Pierenkemper Operating Instructions - MiniFG

General Informations

The device MiniFG is a frequency generator. There are standard programs with fixed
parameters and the possibility of programmable User Programs.

To Use the Device Safely

Please read these Instructions for Use through carefully before using the device!
1.
The MiniFG device must only be used with its original accessories.
2.
Keep the MiniFG away from water or other liquids.
3.
Never use the MiniFG device if it is working incorrectly or has been damaged.
4.
The MiniFG should not be operated at the same time as high frequency devices such
as mobile telephones because there can result in an interaction between the electro-
magnetic fields of such devices with the MiniFG
5.
Do not use the MiniFG in an area where the current will cross the heart.
6.
The used electrodes must not be smaller than 2 cm².
7.
t is only allowed to connect up the MiniFG to one person.
8.
It is not allowed to use the MiniFG during operating machines or during operations
which needs elevated concentration. Pay attention particulary atroad traffic.
9.
Do not drop the MiniFG device, handle it incorrectly, subject it to extreme
temperatures or to a high level of humidity (only use it at temperatures between 10 °C
and 40 °C and at a relative air humidity below 90 %).
10. Care must be taken when the MiniFG basic is used on or in the proximity of children.
11. Store the MiniFG device in its original packaging, to protect it from damage and
contamination.

Contra-Indications

In the following cases the MiniFG should not be used
without prior consultation with the doctor:
In the presence of electronic implants, for example heart pacemakers or pumps
Cardiac arrhythmia
During pregnancy
with seizure disorders (epilepsy)
With skin diseases or wounds in the area of electrode application
With malignant diseases in the area of application

Side Effects

Skin incompatibilities: skin incompatibilities may occur in response to the electrode, the
electrode gel or the current pulses themselves.
In the case of long-lasting redness, burning, itching or skin blisters under the electrodes
or in the region of the electrode system after a stimulation, a consultation with the doc-
16