Compare the value shown with the specifications for the MIG/MAG welding torch cat-
egory on the calibration certificate - see (5) on the image below
-
if the value matches the specifications on the calibration certificate, the testbox is
OK
-
if the value does not match the specifications on the calibration certificate, send
the testbox to the manufacturer to be repaired
Release the flow button on the adapter
Turn the welding torch category switch to the - TIG - position
Press and hold the flow button on the adapter (2)
-
The flow rate is measured
-
the actual value is shown on the testbox display
Compare the value shown with the specifications for the TIG welding torch category
on the calibration certificate - see (6) on the image below
-
if the value matches the specifications on the calibration certificate, the testbox is
OK
-
if the value does not match the specifications on the calibration certificate, send
the testbox to the manufacturer to be repaired
Release the flow button
Lay the adapter hosepack in a straight
1
line
Place the O-ring (3) on the adapter (4)
2
as shown
(3)
Screw the serial number attachment
3
(1) onto the adapter
Start the testbox
4
Turn the welding torch category switch
5
to the - MIG/MAG - position
Press and hold the flow button on the
6
adapter (4)
-
The flow rate is measured
-
the actual value is shown on the
testbox display
(1)