6. Faults

6. Faults
Personnel: skilled personnel
EN
Faults
Contact is no longer
switching in accordance
with the specification.
Switching status remains
unchanged despite
reaching the switch point/
reset point.
No pointer movement
despite change in
pressure.
Pointer movement even
though depressurised.
For claims, the serial and product numbers must be stated. The serial number is printed
on the dial, the product number on the product label. For claims, the atmospheric
pressure and the temperature during the measurement must be given, as well as the
data on the reference standard (model, class).
22
If faults cannot be eliminated by means of the listed measures, the
instrument must be taken out of operation immediately.
Ensure that pressure or signal is no longer present and protect
against accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions given in
chapter 8.2 "Return".
For contact details, see chapter 1 "General information" or the back page
of the operating instructions.
Causes
Electrical connection is
interrupted.
Electrical load unsuitable for
the switch contact model.
Contact contaminated.
Contacts defective (e.g. fused
contact zone).
Movement blocked.
Warming or cooling of the
measuring instrument (no
fault)
Measures
Carry out a continuity test
on the electrical connection
leads.
Maintain the permissible
electrical loads for the switch
contact model.
Replace instrument. Before
recommissioning the
new instrument, provide
a protective circuit for the
contact.
Replace instrument.
Let the instrument settle for 2
hours at 20 °C
[68 °F].