These commands have to be transmitted as ASCII characters, they may be lower or upper key. Each command must use oDh
(Enter) at its end.
Command
Response
PC > HM8115-2
HM8115-2 > PC
Instrument status
*IDN?
HAMEG HM8115-2
VERSION?
version x.xx
STATUS?
function, range
General commands
VAL?
ranges and results
VAS?
ranges
function and
result
Bus commands
FAV0
none
. FAV1
none
Instrument settings
BEEP
none
BEEP0
none
BEEP1
none
Operating modes
WATT
none
VAR
none
VAMP
none
PFAC
none
AUTO:U
none
AUTO: I
none
MA1
value / function
MA0
none
SET:Ux
none
SET:U1
SET:U2
SET:U3
SET:Ix
none
SET:I1
SET:I2
SET:I3
O p e r a t i n g c o n t r o l s a n d d i s p l a y s
Listing of software commands
Text
Instrument identifi cation request
Request for the software version installed Response e.g.: version 1.01
Request for outputting all present instrument settings
functions:
WATT, VAR, VA, PF
voltage ranges:
U1 = 50 V, U2 = 150 V, U3 = 500 V
Current ranges:
I1 = 0,16 A, I2 = 1,6 A, I3 = 16 A
Request for outputting instrument settings and measurement results.
Example of VAr:
U3=225.6E+0 (225.6V in the 500 V range)
I2=0.243E+0 (0.243 A in the 1.6 A range)
VAR=23.3E+0 (Reactive power of 23.3 VAr)
"OF" indicates range overfl ow. In case the command was sent during a
measurement cycle the response will come after its completion.
Selective request for the parameters and the result of FUNCTION.
Example if PF was selected: U3, I2, PF= 0.87E+0.
Disabling of all front panel controls VOLT, AMPERE, FUNCTION
Enabling of all front panel controls VOLT, AMPERE, FUNCTION
Generates a single acoustic signal
Acoustic signal disabled
Acoustic signal enabled
Active power
Reactive power
Apparent power
Power factor PF
AUTORANGE- function voltage enabled
AUTORANGE- function current enabled
Continuous transmission of parameters and results to the PC
Example of PF selected: U3,I2,cos=0.87E+0
"OF" designates overfl ow. Transmission will be continued until ended by MA0.
Ends transmission started with MA1.
Disables autoranging resp. changes the voltage range to „x(Volt)"
Sets 50 V range
Sets 150 V range
Sets 500 V range
Disables autoranging resp. changes the current range to „x(Ampere)"
Sets 0.16 A range
Sets 1.6 A range
Sets 16 A range
37