Arbitrary-netzgerät
Vorschau ausblenden Andere Handbücher für HM8143:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

ABT
Format:
ABT:<list of values>N<number of repetitions>
ABT:tVV.mVmV tVV.mVmV .... Nn or
ABT:tVV.mVmV tVV.mVmV .... Nn
t = time code 0–9, A, B,C, D, E, F; VV.mVmV = 0–30 V
N = end of table character
n = number of repetitions
n =
0 : Continuous repetition
n =
1-255: Waveform is repeated 1-255 times
Function: Programming of the arbitrary waveform function.
The power supply permits creation of a data list
containing up to 1024 voltage values along with
the corresponding time duration values. This list
is transferred in the form of a series of alternating
values for voltages in the range between 0.00 and
30.0 V and codes representing the time duration of
each voltage; at the end of the list, the number of
repetitions is indicated.
How long each voltage appears at the outputs of the
HM8143 is derived from the following table:
0
=
100 μs
h
1
=
1 ms
h
2
=
2 ms
h
3
=
5 ms
h
4
=
10 ms
h
5
=
20 ms
h
6
=
50 ms
h
7
=
100 ms
h
8
=
200 ms
h
9
=
500 ms
h
A
=
1 s
h
B
=
2 s
h
C
=
5 s
h
D
=
10 s
h
E
=
20 s
h
F
=
50 s
h
Example: It is wished to program the following waveform:
1 s
10.00 V
3 s
30.00 V
100 ms
25.67 V
200 us
2.00 V
It is also wished to repeat this sequence 10 times.
The required data table is as follows:
ABT:A10.00_B30.00_A30.00_725.67_002.00 _002.00_N10 or
ABT A10.00_B30.00_A30.00_725.67_002.00_002.00_N10
RUN/STP
Format:
RUN
Function: Starts waveform generation in ARB mode
Format:
STP
Function : Interrupts the arbitrary function while running.
2 x 100 μs 2.00V
100 ms 25.67 V
2 s+1 s 30.00 V
1 s 10.00 V
R e m o t e c o n t r o l