Table 2
Guidance and manufacture's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the
device should assure that it is used in such an environment.
Immunity test
Electrostatic
±15 kV air
discharge (ESD)
±8 kV contact
IEC 61000-4-2
Electrical fast
input/output lines:
transient/burst
±1 kV
IEC 61000-4-4
power supply lines:
±2 kV
Surge
100 kHz repetition
IEC 61000-4-5
frequency
line(s) to earth: ±2 kV
line(s) to line(s): ±1 kV
Voltage dips, short
0%U
interruptions and
At 0°, 45°, 90°, 135°,
voltage variations on
180°,225°,270° and
power supply input
315°
lines
0%U
IEC 61000-4-11
and
70%U
Single phase: at 0°
0% U
(50Hz/60Hz)
30 A/m
magnetic field Power
50Hz/60Hz
frequency
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
Compliance level
±15 kV air
±8 kV contact
power supply lines:
±2 kV
100 kHz repetition
frequency
line(s) to line(s): ±1 kV
; 0.5 cycle
0% U
T
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
; 1 cycle
0% U
T
and
; 25/30 cycles
70% U
T
Single phase: at 0°
; 300 cycle
0% U
T
30 A/m
50Hz/60Hz
17 I GB
environment - guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
; 0.5 cycle
Mains power quality should be that
T
of a typical commercial or hospital
environment.
; 1 cycle
T
; 25/30 cycles
T
;300 cycle
T
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Electromagnetic