7 Commissioning the device
• If spacing is reduced or the dielectric strength of the measuring device cannot
be guaranteed during repair work, perform a test on completion of the work
(e.g. high-voltage test in accordance with the manufacturer's instructions).
• Before commissioning, check whether the data on the nameplate and the
connectionlabel match the requirements of the device. Applicable device for
plants under an obligation to monitoring are labeled appropriate.
• Before commissioning, load the data from the T-DAT into the new transmitter.
Function "T-DAT SAVE/ LOAD" --> Use the selection LOAD.
• If necesary make an empty-pipe/full-pipe adjustment according to the
description in the Operating Instructions.
• Fill in the added form "Storage" and send it to the given adress for archiving
(Data storage in the data base of Endress+Hauser).
b
S-DAT
T-DAT
A0011684
Conversion transmitter Promag 55 compact version to Promag 55 remote version
c
A0014645
e
A0011963
8 Changing the identification
1. After conversion affix the notice label (1) as indicated in the graphic.
a
1
Instrument modified,
for the new Order-Code specifications of the
transmitter, see nameplate on transmitter.
Gerät wurde umgebaut,
für die neue Order-Code-Ausprägungen des
Messumformers ist das Typenschild auf
dem Messumformer zu beachten
Instrument modifié,
pour les nouvelles spécifications du code de
commande du transmetteur, observez
l'etiquette signalétique sur le transmetteur
Date of modification:
Umbaudatum:
Date de modification:
PH2
T20
T20
0.5 × 3.5
22 mm
1
1
H
DN 40...100
1/2
W/P/S
(1 ...4")
A0011961
A0011964
1
H
DN 2...25
(1/12"...4")
A0014774