Conversion transmitter Promag 55 compact version to Promag 55 remote version
2 Safety Instructions
• Comply with national regulations governing mounting, electrical installation,
commissioning, maintenance and repair procedures.
• Requirements with regard to specialized technical staff for the mounting,
electrical installation, commissioning, maintenance and repair of the
measuring devices:
– trained in instrument safety
– familiar with the individual operation conditions of the devices
– for Ex-certified measuring devices: also trained in explosion protection
• Use genuine parts from Endress+Hauser only.
• Using the material number and identification information, check whether the
conversion kit matches the identification label on the measuring device.
• If you have any questions, contact your E+H service organization.
• Follow the documentations for the device.
• Before removing the device: set the process in a safe condition and purge the
pipe of dangerous materials.
• Hot surfaces! Risk of injury! Before commencing work, allow the system and
measuring device to cool down to a touchable temperature.
• In the case of measuring devices in safety-related applications in accordance
with IEC 61508 or IEC 61511: After repair recommission in accordance with
Operating Instructions. Document the repair procedure.
• In the case of Ex-certified measuring devices: Only open in a de-energized
state (once a delay of 10 minutes has elapsed after switching off the power
supply) or in environments which do not have a potentially explosive
atmosphere.
• Service connector:
– do not connect in potentially explosive atmospheres.
– only connect to Endress+Hauser service devices.
4 Reassembly Transmitter Promag 55
"
Caution!
Cut off the power of the device bevore conversion. Only switch on power after conclusion of all works.
a
3 mm
e
T20
Endress+Hauser
b
27
26
27
26
25
24
25
24
23
23
22
22
21
20
21
20
N (L-)
2
N (L-)
2
L1 (L+)
1
L1 (L+)
1
0.5 × 3.5
PH 2
A0011298
f
A0011380
• Only open housing for a brief period. Avoid the penetration of foreign bodies,
moisture or contaminants.
• Replace defective seals/gaskets with genuine parts from Endress+Hauser only.
• Threads (e.g. of the cover for the electronics and connection compartments)
must be lubricated. Use an acid-free, non-hardening grease if an abrasion
resistant dry lubricant is non-existent.
• If threads are damaged or defective, the measuring device must be repaired.
• If spacing is reduced or the dielectric strength of the measuring device cannot
be guaranteed during repair work, perform a test on completion of the work
(e.g. high-voltage test in accordance with the manufacturer's instructions).
• Risk of damaging electronic components! Ensure you have a working
environment protected from electrostatic discharge.
• The measuring device is energized.
Danger: Risk of electric shock! Open the measuring device in a de-energized
state only.
• Observe the instructions for transporting and returning the device outlined in
the Operating Instructions.
3 Preparatory operations: Saving device data
Save the transmitter parameters to the T-DAT using the "T-DAT SAVE/LOAD"
function (Description of device functions or Operating Instructions
"Commissioning" section --> Data backup with "T-DAT SAVE/LOAD).
Replace the T-DAT into the transmitter after conversion.
Overview data storage (Histo-ROM) and F-CHIP
Data storage
Colour
S-DAT
blue
T-DAT
black
F-CHIP (optional)
read
c
22 mm
A0010894
g
S-DAT
T-DAT
A0011381
Proline Promag 55S
Promag 55 compact to remote version
Reassembly
Reassembly
Reassembly
d
A0010895
h
A0011383
3 mm
A0010926
A0014762
2