Herunterladen Inhalt Inhalt Diese Seite drucken

Technical Data - Cs Instruments Fa 515 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für FA 515:
Inhaltsverzeichnis
Verfügbare Sprachen

Verfügbare Sprachen

Measuring range
Type 0699.0515, FA 515
Type 0699.0517, FA 515
Accuracy:
Pressure range:
Power supply:
Output:
Protection class:
EMV:
Operating temperature:
Storage temperature:
Load for analogue output:
Screw-in thread:
Material of housing:
Sensor protection:
Connection:
Response time t95:
**
Remark: The FA515 comes standard with a 4...20mA output in 2-wire technique and an
RS 485 Modbus output. But it can only be used either the
analogue output 4...20 mA or RS 485 Modbus output.
Both outputs can not be used at the same time due to the 4...20 mA
2-wire technology (low power).
Change from 2 wire technolgy to Modbus and back requires the CS Instruments
service software. (Order.No.: 0554 2007)
Sensor ex-factory:
-80...50 °Ctd pressure dew point resp. dew point in °Ctd
0...100 % RH
-20...70 °C
-80...20 °Ctd ≙ 4...20 mA
Other scales on request, for example, -60 ... 30 ° Ctd ≙ 4 ... 20 mA
-20...50 °Ctd ≙ 4...20 mA
typical ± 1 °Ctd von 20...-20 °Ctd
± 2 °Ctd von -50...-20 °Ctd
± 3 °Ctd von -50...-80 °Ctd
-1...50 bar standard
24V VDC (10..30 VDC)
4...20 mA 2-wire technology
**
RS 485 Modbus
IP 65
DIN EN 61326-1
-20...70 °C (ideal 0...50 °C)
-40...80 °C
< 500 Ohm
G 1/2" stainless steel
Optional: UNF 5/8" or NPT ½"
zinc alloy
sinter filter 50 µm stainless steel
M12, 5-pole
< 30 seconds (descending)
< 10 seconds (ascending)
0695 0515 FA515 (-80...20 °Ctd) with 2-wire technology
0695 0517 FA515 (-20...50 °Ctd) with 2-wire technology
13

TECHNICAL DATA

**
FA 515
V1.07
Inhaltsverzeichnis
loading

Inhaltsverzeichnis