Herunterladen Diese Seite drucken

Medela Vario 8 Serviceanleitung Seite 9

Vorschau ausblenden Andere Handbücher für Vario 8:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Sm_177.2610_Booklet_endefr 30.05.2006 14:11 Uhr Seite 5
lX
Battery management
1. Performance of
rechargeable battery
Duration of pumping
Charging time
2. Testing of rechargeable battery
1. Disconnect the mains cord
2. Switch on the Vario
3. Plug the suction port
4. Set required vacuum
5. Measure the running time
6. Compare the running time
ok?
not OK
Conditioning the rechargeable battery
A Completely discharge the rechargeable battery
Disconnect the mains cord
Switch on the Vario
Let the Vario run to a standstill
B Charge battery
Plug in the mains cord
the rechargeable battery is being charged
Wait until the charging indicator goes out
C Test rechargeable battery
ok?
not OK, change rechargeable battery
Storage of rechargeable batteries
In order to keep the self-discharge at a minimum, store the Vario and the replacement
rechargeable batteries at temperatures below 25°C (77°F).
Repeat Steps A and B every 60-90 days.
All manuals and user guides at all-guides.com
low
vacuum
~120 min.
~ 5 hrs.
(charged up to 100%)
low
vacuum
-90 mbar
- 9 kPa
-68 mmHg
> 120 min.
OK
if above values are reached
Do not forget to charge battery before use
OK
If above values are reached
Do not forget to charge battery before use
(only for AC/DC versions)
medium
high
vacuum
vacuum
~ 30 min.
~30 min.
~ 5 hrs.
~ 5 hrs.
medium
high
vacuum
vacuum
-55 kPa
-55 kPa
-413 mmHg
-413 mmHg
> 30 min.
> 30 min.
EN
A1
5

Werbung

loading

Diese Anleitung auch für:

Vario 8 c/iVario 18Vario 18 c/i