6. Commissioning, operation
Confirm selection by pressing ENTER/MENU.
Unit
Now use the arrow keys ▲▼ to select the required calibration function.
Standard linearisation according to DIN IEC 60751 [oFF
Use the arrow keys ▲▼ to select CoFF. Press ENTER/MENU to confirm the required
setting.
Press ESC to return to the measuring mode.
Calibration by code oP1
Use the arrow keys ▲▼ to select oP1. Press ENTER/MENU to confirm the required
setting. A small 1 appears in the lower part of the display. Four characters follow it (hex
code / 0 ... F). The 4 characters can be changed by means of the arrow key up ▲. To go
to the next number press the arrow key down ▼.
If all 4 characters are entered as required, confirm by pressing ENTER/MENU. A small 2
appears and the following 4 characters can be changed as well.
Press ESC to return to the measuring mode.
When calling the oP1 by means of ENTER/MENU the oP1 function
(calibration by code) is activated even if the menu has been closed by
pressing ESC.
WIKA operating instructions, models CTH6300 and CTH6500