7. Faults
Faults
The pointer remains
outside the zero point tol-
erance after installation
and depressurisation.
Instrument outside the
accuracy class.
Vibration of the pointer.
Mechanical damage (e.g.
window, case).
For the replacement of the instrument chapters 9 "Dismounting, return
and disposal" and 6 "Commissioning, operation" must be observed.
WIKA operating instructions pressure gauge, model 2 per ATEX
Causes
Mounting error: Instrument
not mounted in nominal
position.
Transport damage (e.g.
non-permissible shock
loading).
Instrument was operated
outside of permissible
performance limits.
Vibrations in the appli-
cation.
Improper handling.
Measures
Check the mounting
position.
Replace instrument.
Check the observance of
the operating parameters of
the application.
Replace instrument.
Use instrument with case
filling.
Replace instrument.
EN
29

Werbung