Herunterladen Diese Seite drucken

WIKA NS 160 Betriebsanleitung Seite 24

Vorschau ausblenden Andere Handbücher für NS 160:

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Commissioning, operation
For parallel threads, use flat gaskets, lens-type sealing rings or WIKA
profile sealings at the sealing face . With tapered threads (e.g. NPT
threads), sealing is made in the threads , using a suitable sealing
EN
material (EN 837-2).
The torque depends on the sealing used. In order to orientate the
measuring instrument so that it can be read as well as possible, a
connection with LH-RH union or union nut should be used.
When a blow-out device is fitted to a pressure gauge, it must be protect-
ed against being blocked by debris and dirt.
6.2 Requirements for the installation point
If the line to the measuring instrument is not adequately stable, an instru-
ment bracket should be used for fastening (and possibly via a flexible
capillary). If vibrations cannot be avoided by means of suitable instal-
lation, instruments with liquid filling should be used. The instruments
should be protected against coarse dirt and wide fluctuations in ambient
temperature.
24
Spanner flats
Sealing face
WIKA operating instructions pressure gauge, model 2 per ATEX
Sealing in the thread

Werbung

loading

Diese Anleitung auch für:

Ns 100