Herunterladen Diese Seite drucken

KERN MCD Serie Betriebsanleitung Seite 46

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10 Error messages
Display
Description
Zero range exceeded
(on start-up or when pressing the
• Load on weighing pan
• Excess load, during zero setting of weighing scale
• Incorrect adjusting process
• Fault on load cell
Value outside the A/D converter range
• Damaged weighing cell
• Damaged electronics
Unable to initialise zero point
OL or --------
Overload
------or „0"
Underload
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
11 Servicing, maintenance, disposal
11.1 Cleaning
Before any maintenance, cleaning and repair work disconnect the
appliance from the operating voltage.
20
Measuring cell defective / overloaded
Object on weighing pan / contact
Transport safety device has not been removed
Main board defective
Unload, switch off and adjust the balance
Unload, switch off and adjust the balance
key)
MCD-BA-e-1710

Werbung

loading

Diese Anleitung auch für:

Mcd 300k-1