When the „Report" button is activated, the NANOCOLOR
T-Set software creates a certificate of the test run, that is
then automatically printed.

3.6 Calibration

After completing a calibration run, the calibration data
measured by the NANOCOLOR
heating block, can be read out, displayed and edited.
After calibration, information can be entered in the entry
boxes (date of the calibration, number of the NANOCO-
LOR
®
date of issue). In the „Notes box", any notes that may be
relevant can be entered.
Note:
If the heating block does not contain any calibration data
or the data has already been entered, it is not possible to
input any data or amend any existing data (the fields are
greyed). The input of data or any modification is possible,
only after a new calibration run has been completed.
The temperature table is divided into three columns. In
the „nominal" column, all temperatures are listed that have
been programmed in the heating block. The „actual" co-
lumn shows the temperatures measured with the NANO-
COLOR
ted temperatures are listed.
When the „Report" button is activated, the NANOCOLOR
T-Set software creates a certificate of the calibration run,
that is then automatically printed.
18
T-Set.
®
T-Set, the certificate number applicable and the
T-Set. In the „Calibrated to" column, all calibra-
®
T-Set and saved in the
®
®
®