ning. Alternatively, the temperatures can be displayed and documented with the
T-Set software after the test program is finished.
After testing, the data can be transferred to a connected USB stick via the
icon. Via the
you can reset the calibration of the heating block to the factory settings.
Safety Notice:
The temperature sensor is hot!
When removing the sensor from the heating block, grasp the black insulation.

2.2.2 Calibration

The calibration can be started by pressing the
peratures in the heating block are measured with the NANOCOLOR
registered and saved in the heating block (duration approx. 40 min). If the heating
block has reached one of the required temperatures you will hear an acoustic
signal. The temperature which is currently measured by the NANOCOLOR
T-Set is displayed.
The temperature which is measured by the NANOCOLOR
documented manually while the test is running. Alternatively, the temperatures
can be displayed and documented with the T-Set software after the test program
is finished.
When all temperatures which are programmed in the heating block are calibrated,
the
icon appears. After confirming the
new calibration should be saved. Confirm with the
new calibration or press the
calibration process is finished the data can be transferred to a connected USB
stick via the
With the
icon you can reset the calibration of the heating block to the
factory settings.
Safety Notice:
The temperature sensor is hot!
When removing the sensor from the heating block, grasp the black insulation.

2.2.3 External measurements

Besides the temperature control of the heating blocks, the NANOCOLOR
T-Set can also be used for external temperature measurements. For this purpose,
plug the NANOCOLOR
temperature is now shown on the display of the NANOCOLOR
icon you get back to the superior menu. With the
icon to reject the new calibration. When the
icon. Via the
USB-T-Set into an USB port of your choice. The current
®
NANOCOLOR
®
icon. All programmed tem-
icon you are asked, if the
icon you can get to the superior menu.