Maintenance

Sensor cap
Sensor holder
8.2

Sensor exchange

9

Accessoires

The service life of the sensor is about 1 year in
normal use. It is shortened with constant use
when the refrigerant concentration is high. Proper
maintenance can extend the life of the sensor.
Switch off the unit.
Unscrew the sensor protection cap and wash
and dry it.
Blow impurities off the sensor with dry air.
Store the unit in a dry and clean place and re-
move the batteries if you are not going to use
it for a long time.
If the sensor is defective, error message 2E ap-
pears in the display. In this case, replace the sen-
sor as follows:
Unscrew the sensor cap.
Pull the sensor out of the holder by hand or
with needle-nose pliers.
Place the new sensor in the holder, making
sure to push the contacts of the sensor into
the openings of the sensor holder.
Replacement Sensor Wöhler RL 200:
Art. No.13028