Model:
FLXA21 / FLXA202
Installation for Zone 2 / Division 2
Applicable models: FLXA21-D-x-x-CD-xx-xx-F-..., FLXA21-D-x-x-CD-xx-xx-P-...
FLXA21-D-x-x-DD-xx-xx-F-..., FLXA21-D-x-x-DD-xx-xx-P-...
Hazardous Area
Class I, Zone 2, Group IIC, or
Class I, Division 2, Groups A, B, C, D
Temperature Class: T4
FLXA21 Analyzer
Housing Assembly
Measuring Module 1
Sensor 1
(Note 6)
Measuring Module 1 (Note 6):
Type of Measuring Module
pH, SC, DO
Uo
11.76 V
Io
116.5 mA
Po
0.3424 W
Co
100 nF
Lo
1.7 mH
Specific condition of use
-
Electrostatic charges on the non-metallic or coated parts of the two wire analyzer shall be
avoided.
Rev.
Other
Field Devices
Supply +
T
Supply –
Terminators
Supply +, Supply – (Note 7):
Ui: 24 V
Ci: 2.72 nF
Li: 0 mH
ISC
SENCOM
11.76 V
5.36 V
60.6 mA
106.16 mA
0.178 W
0.1423 W
100 nF
31 μF
8 mH
0.45 mH