Herunterladen Inhalt Inhalt Diese Seite drucken

Characteristics; Reference Conditions; Electrical Characteristics, Sensitivity 1Mv/A - Chauvin Arnoux PAC 25 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 42

3.1. REFERENCE CONDITIONS

Quantity of influence
Temperature
Relative humidity
Position of the conductor
Frequency of the measured signal
External electric field
External DC magnetic field (earth's field)
External AC magnetic field
Impedance of the measuring instrument
The intrinsic uncertainty is the error defined under the reference conditions.
It is expressed as a percentage of the output signal (R) plus an offset in mV:
± (a % R + b)

3.2. ELECTRICAL CHARACTERISTICS, SENSITIVITY 1mV/A

Output impedance: 215 W
Specified
measurement
0.5 to 100 Aac/dc
range
Intrinsic
≤ ± (2 %L + 1.5 mV)
uncertainty
Phase error (45 to 65 Hz)
Specified measurement range
Phase shift

3. CHARACTERISTICS

100 to 800 Aac/dc
≤ ± 2.5 %L
3 to 200 Aac
≤ ‑2°
Reference values
centred on the marks on the clamp
DC to 65 Hz sine wave
≥ 1 MW and ≤ 100 pF
800 to 1000 Aac/dc
≤ ± 4 %L
200 to 1000 Aac
≤ ‑1.5°
30
23 ± 5°C
20 to 75 %RH
zero
< 40 A/m
zero
1000 to 1400 Adc
≤ ± 5 %L

Werbung

Inhaltsverzeichnis