Limit the measuring time between 1000 and 1200 A : 15 minutes on, followed by 30 minutes off.
With the C100 model, which is not protected by an output limiter, do not clamp a conductor before
connecting the clamp to the corresponding measuring device. Likewise, do not disconnect the clamp
from the measuring appliance when the clamp is still attached to the cable.
Before connecting the clamp to the multimeter, check that the multimeter has an appropriate calibre.
Open the jaws and clamp the cable through which the current you wish to measure is running.
Roughly centre the cable in the jaws. Follow the direction of the arrow, if so required by the
application in question.
To read the measurement, apply the appropriate reading coefficient (See 4.2 "Input/output ratio").
NB : for multi-calibre models, select the ratio which provides the best resolution and precision.
4.1

Reference conditions

Temperature
Humidity
Conductor centred in jaws
Sinusoidal current
Distortion factor
Direct current
Continuous magnetic field
Alternating magnetic field
Proximity of external conductors : no current
Measuring device impedance
- C100 / C102 / C103 : ≤ 5 Ω
- C112 / 113
- C122 / C148
4.2

Specifications and references for ordering

Reference
Model
for
ordering
Alternating current output
P01.1203.01
C100
P01.1203.02
C102
P01.1203.14
C112
P01.1203.03
C103
P01.1203.15
C113
P01.1203.06
C122
P01.1203.07
C148

3. USE

4. CHARACTERISTICS

: +20... +26°C
: 20... 75% RH
: 48... 65 Hz
: < 1%
: no
: earth field (< 40 A/m).
: no
- C106 / C107 / C116 / C117 : ≥ 1 MΩ and ≤ 100 pF
: ≤ 1 Ω
- C160
: ≤ 0.2 Ω
- C173
Nominal
measuring
scope
100 mA... 1000 A
10 mA... 1000 A
100 mA... 1000 A
10 mA... 1000 A
1 A... 1000 A
1 A... 1000 A
11
: ≥ 1 MΩ and ≤ 47 pF
: ≥ 10 MΩ and ≤ 100 pF
Input/Ouput
Connec-
ratio
tion
Sockets
1 mA~ / 1 A~
Sockets
1 mA~ / 1 A~
Leads
5 mA~ / 1 A~
Sockets
5 mA~ / 1 A~
10 mA~ / 1 A~
Sockets
20 mA~ / 1 A~
Ouput
protected
against
voltage surges
No
30 V
max. peak