Herunterladen Inhalt Inhalt Diese Seite drucken

Calling The Quick Menu; Opening Information; Opening Device Information - Dräger X-am 2800 Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für X-am 2800:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 3
en
|
Operation
If the measuring range is exceeded significantly at the CatEx channel (very high
concentration of flammable substances), a blocking alarm is triggered. This
CatEx blocking alarm can be acknowledged by switching the gas detector off
and then on again in fresh air.
If the gas detector cannot be switched off because the A2 alarm is active and
the switch-off mode in the CC-Vision is set to "Switching off not allowed during
A2", remove the power pack or place the gas detector in the charging cradle and
allow it to switch off automatically.
For more details, refer to the technical manual.
4.7

Calling the Quick menu

The Dräger CC-Vision software can be used to save preferred features in the
Quick menu.
Possible features:
Fresh air calibration
Bump test
Delete peak values
Bluetooth
Show pump information (only when pump adapter connected)
Activate or deactivate pump (only when pump adapter connected)
To open the Quick menu:
1. Press
3x in measuring mode.
2. Press
to scroll through the available features.
3. Press
to call the selected feature.
OK

4.8 Opening information

4.8.1 Opening device information

This feature can be used to call up device, channel, firmware and Bluetooth
Label) information and peak evaluations.
If any warning or fault messages exist, the corresponding note and fault codes
are displayed. Further information on the individual fault codes for service and
maintenance is provided in the technical manual.
If no key is pressed for 10 s, the gas detector automatically returns to
measuring mode.
54
®
connection
®
X-am
2800 / X-am
®
(e-
®
5800

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

X-am 5800

Inhaltsverzeichnis