Declaration – Electromagnetic immunity
EN
This smart blood pressure monitor with ECG & digital stethoscope system is intended for use in the electromagnetic environment specified below. The customer or the user of
FR
the smart blood pressure monitor with ECG & digital stethoscope system should ensure that it is used in such an environment.
DE
SV
Immunity test
IEC 60601 test level
NL
FI
Electrostatic
Contact: ±8 kV
DA
discharge (ESD)
Air: ±2 kV, ±4 kV, ±8
IEC 61000-4-2
kV, ±15 kV
IT
ES
Electrical fast
2 kV for power
CS
transient/burst
supply lines
PL
IEC 61000-4-4
1 kV for input/output
lines
PT
RO
Surge IEC
1 kV differential
61000-4-5
mode
HU
2 kV common mode
SK
ET
Voltage dips,
-5% UT(95% dip in
short interruptions
UT) for 0.5 cycle,
EL
and
-40% UT (60% dip in
IS
voltage variations
UT) for 5 cycles,
on power
-70% UT (30% dip
LV
supply input lines
in UT) for 25 cycles,
LT
IEC 61000-4-11
-5% UT (95% dip in
SL
UT) for 5 sec
TR
Power frequency
30 A/m
BG
(50/60 Hz)
50 Hz or 60 Hz
magnetic field IEC
HR
61000-4-8
Compliance
Electromagnetic
level
environment – guidance
Contact: ±8
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity
kV Air: ±2 kV,
should be at least 30%.
±4 kV, ±8 kV,
±15 kV
N/A
The main power quality should be similar to that of a typical commercial or hospital environment.
N/A
N/A
The main power quality should be similar to that of a typical commercial or hospital environment.
N/A
N/A
The main power quality should be similar to that of a typical commercial or hospital environment. If the user of the equip-
ment or system requires continued operation during power main interruptions, it is recommended that the equipment or
system be powered from an uninterruptible power supply or a battery.
30 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital
50 Hz or
environment.
60 Hz