6 Technical Data and Operating Limits

6.1 Maximum Verification Scale Interval

Max. verification scale
interval [e] / OIML
1 x 3000e Max / e [kg]
Max. verification scale
interval [d] / NTEP
1 x 5000d Max / d [lb]

6.2 Maximum Permissible Load

All weighing platforms are equipped with an overload
protection. With the robust scale design you can
occasionally exceed the rated capacity of the platform
without damage. The maximum static safe load must
never be exceeded.
Observe the following operation limits.
Model
PBA439(x/xx)-QA
PBA439(x/xx)-A
PBA439(x/xx)-QB
PBA439(x/xx)-BB
PBA439(x/xx)-B
PBA439(x/xx)-BC
PBA439(x/xx)-CC
10
Technical Data and Operating Limits​ ​
3 kg
6 kg
15 kg
3 / 0.001
6 / 0.002
15 / 0.005 30 / 0.01
-
10 lb
25 lb
-
10 / 0.002 25 / 0.005 50 / 0.01
A - Central Load
40 kg / 80 lb
40 kg / 80 lb
100 kg / 200 lb
100 kg / 200 lb
200 kg / 400 lb
400 kg / 800 lb
700 kg / 1400 lb
WARNING
Avoid falling loads, shock loads as well as impacts from the side.
Capacity
30 kg
60 kg
150 kg
60 / 0.02
150 / 0.05 300 / 0.1
Capacity
50 lb
100 lb
250 lb
100 / 0.02 250 / 0.05 500 / 0.1
B - Side Load
C - Corner Load
30 kg / 60 lb
15 kg / 30 lb
30 kg / 60 lb
15 kg / 30 lb
70 kg / 140 lb
35 kg / 70 lb
70 kg / 140 lb
35 kg / 70 lb
140 kg / 280 lb
75 kg / 150 lb
300 kg / 600 lb
150 kg / 300 lb
400 kg / 800 lb
200 kg / 400 lb
300 kg
600 kg
600 / 0.2
500 lb
1000 lb
1000 / 0.2

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis