CAUTION!
Take care not to touch the surface of the glass. Ideally wear gloves, handle with soft cloth and wipe the surface
with alcohol before fitting into the UV lamp housing.
7. Unpack the new UV Lamp.
9. Insert the cap and tighten the screws.
10. Connect the lamp connector to the end of the UV.
6. Lift and remove the UV lamp from the housing.
Discard the used UV lamp accordance
with local regulations.
(Re-use the red O ring on the new UV lamp)
Chorus 1 Operator Manual MANU39998 LSC ANR GSC Version 02 3/17 Page 9
8. Slide the new UV Lamp into the housing.
11. Turn on the water supply.
12. Turn on the power supply.