Display
What it means
Insufficient sample
The meter does not
begin to test after
Sample is applied to wrong
applying sample
site of the test strip
Used test strip
The test strip is exposed in air
too long after taking out from
Meter displays
the vial
incorrect test result
Press the strip too firmly
against your finger, sample
cannot be smoothly applied.
The test strip foil pouch is
broken or damaged or left
open to air. The test strips in
vial is beyond the expiration
or the discard date.
47
What to do
Repeat the test with a new test strip
and enough blood sample
Apply the sample into test strip according
to the instruction of "Applying sample"
Use new test strip
Finish the test within 3 minutes after
taking out the strips from the vial
Gently touch the blood drop to the top
edge of the test strip.
Use new test strip
Display
What it means
Expired test strip
Meter displays
Defective meter or test strip
incorrect test result
false test mode
The test strip is exposed in
moisture environment for too
without sample
long
applied to, test starts
and meter counts
down as soon as
The test strip is exposed in
the test strip is
moisture environment for too
inserted in
long
If you follow the actions recommended but the problem is not resolved, please contact Customer
Service at +86-731-89935581/+86-731-89935582 or contact your local dealers for assistance.
What to do
Use new test strip
Contact the manufacturer or the dealer
choose the right test mode
use a new test strip
use a new test strip
48