STEP 9: Insert strip into the test strip port, The
AQ UG meter will beep and turn on.
The port marked with "G" is for blood glucose strip and
the port marked with "U" is for uric acid strip. If
blood glucose strip inserted, "GLU" will be displayed
on the bottom-left of the screen, if uric acid strip inserted,
"UA" will be displayed on the top-left of the screen.
After test strip inserted, the meter will display flashing
blood drop symbol and you can apply blood sample.
Attention:
If blood is not applied within 3 minutes, the meter will automatically turn off. Reinsert the
strip and begin test.
If the meter display "E-2", it means the environment temperature is beyond 40℃ or
below 5℃, Please put the meter, strip, control solution at appropriate place (10℃~35℃),
wait at least 30 minutes and test again.
The test strip port must be match the inserted strip, or else the meter will not turn on.
The meter can test blood glucose and uric acid together.
25
Safe
STEP 10:Take Blood Sample
Make sure the lancing device is pressed firmly to
your finger. Press the button to lance.
Attention: If you didn't get enough blood sample,
please gently massage your finger. DO NOT
SQUEEZE YOUR FINGER. Refer to step 6 to
adjust puncture depth and lance again.
STEP 11: Apply blood sample
When the meter screen displays a flashing blood drop
symbol "
", apply the blood sample to the reaction
chamber.
Attention:
Apply blood sample only the blood drop symbol "
Before applying blood sample, the disinfection ethanol in the puncture site should
completely volatilize.
Hold the tip of the test strip to the blood drop until the meter beeps. Visually confirm that
the blood sample fills the entire reaction chamber at the end of the test strip.
Do not move meter or strip and press main button when testing.
If you didn't apply enough blood sample to test strip, use a new test strip to test again.
Do not use test strip press finger firmly, otherwise it may cause inaccurate test result.
" flashing.