Questionable Results:
Probable Cause
Operation mistake
Control solution not being shaken well
Use the first drop of control solution
Expired or contaminated control solution
Expired or contaminated test strip
Temperature is too high or too low
Wrong correcting code
Meter malfunction
21
Action
Retest according to correct steps
Shake control solution well and retest with a new test strip.
Retest with a new test strip
Change a new vial control solution and retest.
Change a new vial test strip and retest.
Equilibrate meter, test strip and control solution in room
temperature for 30 minutes and retest.
Set correcting code which accordance with test strip
package and retest.
Contact customer service or local dealers

Blood Glucose and Uric Acid Testing

Prepare meter, strip, lancing device and lancet before test.
STEP 1: Wash your hands or use an alcohol swab to
clean the fingertip. MAKE SURE YOUR HAND IS
DRY BEFORE TESTING.
STEP 2: Unscrew the lancing device cover.
STEP 3: Insert a lancet into the lancing device until
it comes to a full stop.