Sinocare Safe AQ UG Benutzerhandbuch Seite 26

Blutzucker- und harnsäuremonitoringsystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Display
What it means
Your test result is higher than
20.0 mg/dL or 1188μmol/L.
Low battery power.
Meter is out of the required
testing temperature range.
43
What to do
Use a new test strip to test again, if
meter still display "HI", call healthcare
professional immediately.
Replace the batteries immediately.
Place the meter and test strip at the
operating temperature range.
Repeat the test when the meter and
test strips have reached a
temperature within the operating
range.
Display
What it means
For blood glucose testing
Test method error: insert strip
into meter after apply sample
to strip.
Used blood glucose test strip
Expired test strip or damped
test strip
Test strip exposed to air more
than 3 minutes.
For uric acid testing
Test method error: insert strip
into meter after apply sample
to strip.
Used test strip
What to do
Insert test strip into meter before
apply sample to strip.
Use a new test strip to test again.
Use a new test strip to test again.
Use a new test strip to test again.
Insert test strip into meter before
apply sample to strip.
Use a new test strip to test again.
44