Display
What it means
Expired test strip or damped
test strip
Uric acid coding failure
Parameter error.
45
What to do
Use a new test strip to test again.
Coding again
Call Customer Service at +86-731-
89935581/+86-731-89935582 or
contact your local dealers.
Display
What it means
Battery incorrectly installed
Battery power is low
Test strip inserted incorrectly
There is no feedback on the
Meter does not turn
display
on after inserting a
test strip
Test strip Inserts into wrong
ports
Poor contact with Insert port
What to do
Check that the battery is correctly
installed. Pay attention to the "+"sign
Replace battery
Insert the test strips with the arrow
side facing up until it will go no further
Remove the battery and wait at least 3
minutes, then reinstall the battery. If
the malfunction has not been solved,
please contact the manufacturer or
the dealer.
Make sure the right port when insert
the test strips.
The blood glucose test strip insert
into port marked with "G".
The uric acid test strip insert into
port marked with "U"
Please contact the manufacturer or
the dealer.