color2view
4 Measurement Parameter
Go to Configuration > Measurement parameter.
Traceable to
Illumination
Aperture
Observer
Color system
Fluorescence
Color equation
Statistics
Color indices
Gloss
Always use auto-
standard
Always continue
last test series
8
Select the measurement mode:
• NRC: Results can be compared with results of the
"National Research of
• Color-view: Results can be compared with the
"color-view", the predecessor of the "color-2-
view".
Default is NRC.
Select standard illuminant. Default is D65.
Select aperture mounted on the instrument. Default is
32 mm.
Select standard observer. Default is 10°.
Select color system. Default is CIE L*a*b*.
Refer to
"Fluorescence Measurement
ΔFl.
Select color equation. Default is ΔE*.
No. of readings to be taken per sample. Statistics are
evaluated, if n > 1.
Select color indices.
Default is none.
Turn gloss measurement on or off.
Search for nearest standard is always active.
No new test series is created; latest series is opened
automatically.
BYK-Gardner GmbH
Canada".
[} 13]". Default is