13. NTU-Check

Plus
The PF-12
enables a warning of interfering turbidity when measuring a test kit. To activate/
deactivate the NTU-Check, enter 958 as method in the start screen. Use the keys
deselect the check and confirm your selection with
Enter the desired warning limit using the keypad and confirm with
When this function is activated, a nephelometric turbidity measurement is performed
parallel to each measurement. MACHEREY-NAGEL recommends the use of 10 NTU
as a warning limit.

14. Quality control

Each institution must implement its own QC policy.
To check the performance of the test kit and instrument used, we recommend measuring
several standard solutions with known concentrations within the measuring range.
The photometric accuracy of the measurement of the PF-12
the NANOCONTROL NANOCHECK 2.0 color solutions (REF 925703) provided by
MACHEREY-NAGEL.
Manual Photometer PF-12
60
Tube for correction (value B)
Fill empty test tube with 0.2 mL R2, 4.0 mL sample solution and 1.0 mL
distilled water, close, mix. Analytical result = A + B
Fill empty test tube with 4.0 mL sample.
Fill empty test tube with 4.0 mL sample, add 0.2 mL R2, close and mix.
Proceed as described in the test instruction for test 0-97,
add 1.0 mL ethanol instead of R4.
Proceed as described in the test instruction for test 0-98,
add 0.5 mL distilled water instead of R3.
Open ethanol test tube, add 4.0 mL R1 and 0.5 mL sample solution (consider
dilution), mix, add 2 drops R3, close and mix.
Open methanol test tube, add 3.0 mL R1 and 1.5 mL sample solution
(consider dilution), mix, add 2 drops R3, close and mix.
Fill empty test tube with 4.0 mL sample.
No correction possible.