Immunity test levels
Test
Electrostatic discharge
(IEC 61000-4-2)
Radiated RF EM fi elds
(IEC 61000-4-3)
Proximity fi elds from RF wireless
communications equipment
(IEC 61000-4-3)
Electrical fast transients and
bursts
(IEC 61000-4-4)
Surge Immunity (IEC 61000-4-5)
Conducted disturbances induced
by RF fi elds (IEC 61000-4-6)
Rated power frequency magnetic
fi elds (IEC 61000-4-8)
13.2. Operating principle
The Bloom Lab is an electronic device, which contains an optical measurement system (optical reader) to quantify the test
line intensities produced by the Bloom Test strip during use. It captures the Bloom Test strip read-out window, automatically
detects the test and control line and calculates the mean intensity value for each line by image analysis. This mean value,
in connection with further image analysis, corresponds to the concentration of the captured target analyte. The actual
concentration of the target analyte is calculated diff erently for each batch of Bloom Tests using the Bloom App, thus enabling
measurements with the Bloom Lab for a wide variety of analytes.
4-3-6-2 Bloom Lab IFU, 09.04.2020, V1.0
Test levels
Contact Discharge: ±8 kV
Air Discharge: ±2 kV, ±4 kV, ±8 kV, ±15 kV
80-2700 MHz; 1kHz AM 80 %; 10 V/m
385 MHz; Pulse Modulation: 18 Hz; 27 V/m
450 MHz; FM: ±5kHz Hub, 1 kHz Sinus; 28 V/m
710, 745, 780 MHz; Pulse Modulation: 217 Hz; 9 V/m
810, 870, 930 MHz; Pulse Modulation: 18 Hz; 28 V/m
1720, 1845, 1970 MHz; Pulse Modulation: 217 Hz; 28 V/m
2450 MHz; Pulse Modulation: 217 Hz; 28 V/m;
5240, 5500, 5785 MHz; Pulse Modulation: 217 Hz; 9 V/m
Power supply lines: ±2 kV, 100kHz repetition freq.
Signal / Data lines: ±1 kV, 100kHz repetition freq.
0.5, 1,kV; Diff erential mode
Surge Waveform: Combination Wave
(2 µs x 50 µs Voltage, 8 µs x 20 µs Current)
0.15-80 MHz; 1kHz AM 80 %; 3 Vrms, 6 Vrms ISM band
30 A/m, 50 Hz / 60 Hz
Table N°5: EMI test levels
EN
13