Transmitter
(2)
Field shape
Elevation
The Transmitter and Receiver should be positioned at least 5  feet (1.5  m) above the floor, be at the same
height, and, preferably, with the Transmitter and Receiver front edges parallel and facing each other.
Horizontal distance between Transmitter and Receiver
ZEROWIRE® G2 operates correctly with a horizontal distance between Transmitter and Receiver of up to
30 feet (9.1 m). However, in the case of most OR environments, best results are achieved with a horizontal dis-
tance of up to 8 feet (2.4 m).
(2)
Set up and Installation