1 OVERVIEW
Incorporating a flat sensor, the Elcometer 138 Bresle Salt Meter has been designed primarily for use in
accordance with ISO 8502-6 / ISO8502-9. It measures the conductivity of the sample but displays the surface
density of salts. The measured conductivity is converted to surface density of salts using the ISO Salt Mix or
a
IMO PSPC equivalent NaCI conversion factors below, dependent on the measurement mode selected, based
2
on an area of 12.5cm and a sample volume of 3ml.
For detailed information on taking measurements in accordance with ISO 8502-6 /ISO8502-9, please refer to the
user guides for the Elcometer 138 Bresle Salt Kits.
Reading
µS/cm
The Elcometer 138 Bresle Salt Meter can also be used in accordance with US NAVY PPI 63101-000 (Rev 27),
ISO 8502-11; AS 3894.6-A and SSPC Guide 15.
2 BOX CONTENTS
Elcometer 138 Bresle Salt Meter & Sensor
<
Conditioning Solution; 14ml (0.47 fl oz)
<
CR2032 Lithium Batteries; x2
<
User Guide
<
Note: The Elcometer 138 Bresle Salt Meter measures aqueous solutions. The meter is NOT designed to measure solids, organic
solvents, surfactant, oil, adhesive, alcohol, strong acids (pH: 0 to 2) or strong alkalis (pH: 12 to 14). The life of the sensor
will be extremely short if these substances are measured.
a
International Maritime Organisation, Performance Standard for Protective Coatings.
b
2
Based on an area of 12.5cm and a volume of 3ml.
www.elcometer.com
Surface Density of Salts: Factors
ISO Salt Mix
2
mg/m
µg/cm
x1.2
x0.12
(supplied fitted to the Elcometer 138)
IMO PSPC equivalent NaCI
2
2
mg/m
x1.1
b
2
µg/cm
x0.11
R