5.
Make sure that the error is in the specified limits (Table 8).
Applied
mA
±55
±40
±24
±18
±12
±6
0 (open circuit)

4.4 Procedures (IDOS UMM)

Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument automatically sets a new calibration date in the
UMM.

5. Specification Data

All accuracy statements are for one year.

5.1 General

Item
Languages
Operating temperature
Storage temperature
Humidity
Shock/Vibration
EMC
Safety
Size (L: W: H)
Weight
Power supply
Duration
(Measure)
Copyright 2003 Baker Hughes Company.
22 | DPI 800/802 Instruction Manual–English
Table 8: mA Input Error Limits
Calibrator
Error
(mA)
0.0022
0.0018
0.0014
0.0004
0.0003
0.0002
n/a
Specification
English [Default]
14 ... 122°F (-10 ... 50°C)
-4 ... 158°F (-20 ... 70°C)
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
EN 61010; Def Stan 66-31, 8.4 cat III
EN 61326-1
Pressure Equipment Directive - Class: Sound Engineering Practice
(SEP)
CE Marked
Maximum: 7.5 x 3.3 x 2.0 in
(190 x 85 x 50 mm)
19 oz (530 g)
3 x AA alkaline batteries
P1: ≈ 50 hours
P1 + mA: ≈ 5 hours (24 V Source at 12 mA)
Permitted
DPI 800/802 Error
(mA)
0.005
0.004
0.003
0.003
0.002
0.002
0.001

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Druck dpi 802

Inhaltsverzeichnis