Before you disconnect the instrument, isolate and bleed the system.
P1
Figure 2: Example Configuration - Measure pressure
The examples in Figure 2 shows:
How to measure one or two pressures in a system.
How to measure pressure difference.
Note: ∆P = High range IDOS - Low range IDOS. If Pressure (p2) > Pressure (p1), the result is
negative.

2.4 Leak Test

To do a leak test on a pressure system:
1.
Connect the instrument (Figure 3) and, if necessary, adjust the Set Up (Table 1).
2.
Select an applicable leak test from Select Task (Table 2/Table 3) and, if necessary, adjust
the Settings (Table 4).
3.
Set the period for the leak test (Table 4).
4.
If necessary, do a zero correction (Table 4).
5.
To start the leak test, select Start (). When the test is finished, the instrument calculates
the leak rate in the applicable units/minute.
Copyright 2003 Baker Hughes Company.
12 | DPI 800/802 Instruction Manual–English
p1
p2
P2
P1 ... g
P
MENU
ESC
OK
HOLD
a) P1 and/or P2
p1
p2
P1 30 ... g
MENU
OK
30 ... g
15.006
20 ... g
10.003
Settings
P
MENU
OK
P1-P2
5.003
Settings
b) P1 - P2