Before you disconnect the instrument, isolate and bleed the system.
+
-

2.6 Transmitter Calibration

DPI 802/802P only. To calibrate a transmitter:
1.
Connect the instrument (Figure 5 or Figure 6) and, if necessary, adjust the Set Up (Table 1).
2.
Select the applicable calibration task from Select Task (Table 2/Table 3) and, if necessary,
adjust the Settings (Table 4).
3.
If necessary, do a zero correction (Table 4).
4.
For each calibration point, apply an applicable pressure and wait for the pressure system to
stabilize.
Copyright 2003 Baker Hughes Company.
14 | DPI 800/802 Instruction Manual–English
Figure 4: Example Configuration - Switch Test
P1
MENU
ESC
OK
HOLD
mA
24V
30V MAX
30V MAX
P1
MENU
OK
30 ... g
5.003
7.45
6.27
Settings

Werbung