Herunterladen Inhalt Inhalt Diese Seite drucken

Error Messages - Mettler Toledo ICS226 Bedienungsanleitung

Kompaktwaage
Vorschau ausblenden Andere Handbücher für ICS226:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 21
Error code
Error
Overload
Underload
Zero/tare setting outside setting
range
Key function cannot be performed
Scale is in motion
EEPROM checksum error
Calibration weight too small
Scale not calibrated yet
Illegal operation
Unit error in OIML mode
18

Error messages

All manuals and user guides at all-guides.com
5 Error messages
Remedy
➜ Unload scale.
➜ Reduce preload.
➜ Set the scale to zero.
➜ Unload scale.
➜ Empty load plate before power up.
➜ Recalibrate zero point.
➜ Go back to gross mode.
➜ Ensure the scale is stable.
➜ Unplug power plug and plug in again.
➜ If the problem persists, contact your authorized METTLER
➜ Check calibration weight.
➜ Recalibrate scale.
➜ Switch off the scale and switch on again.
➜ Change unit to kg or g.
TOLEDO Service.
ICS226 30232252A

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis