Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning And Maintenance; Where To Store The Nebulizer; Troubleshooting - Stocker Geyser Benutzerhandbuch

Nebulisator 4 l und 12 l li-ion
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

The nebulizer must be cleaned after each cycle to prevent the liquids used from causing
corrosion and blocking of the nozzles.
Storage
The most important thing to keep in mind before storing your nebulizer is to make sure that the
battery is fully charged and the power is off.
1.
Make sure that the battery is fully charged.
2.
Make sure that the power is off.
3.
Store the nebulizer in a cool and dry environment with a room temperature from ranging
between 15° and 35° C.
4.
If the nebulizer will not be used for a long time, charge the battery at least once every six
months.
5.
The nebulizer and the battery charger must be stored safely, protected from heat, dust and
moisture.
6.
The nebulizer and flammable substances must be kept at least 1.5 meters apart.
7.
Do not turn on the power switch during storage.
66
Art. 410 / Art. 411
Geyser Nebulizer 4 L and 12 L Li-Ion

Cleaning and maintenance

Where to store the nebulizer

Art. 410 / Art. 411
Geyser Nebulizer 4 L and 12 L Li-Ion

Troubleshooting

Problem
Cause
1. The filter is clogged
The nebulizer does not
2. Faulty motor
spray even if the motor
3. The delivery tube
is running
between the tank and
motor is crushed
1. Some wires have come
loose
The motor does not run
2. Faulty motor
even with the nebulizer
3. The pressure is too
switched on (the LED
high or too low
above the start key is
4. The tank is empty
lit red)
5. The spraying intervals
are set incorrectly.
1. The battery is dead
2. Faulty battery
The nebulizer does not
3. Faulty PCB/display
start (the LED above the
start key is off)
4. Some wires have come
loose
1. Faulty PCB/display
2. The battery is dead
The display is not
3. Faulty battery
working
4. Some wires have come
loose
The keys do not work
1. Faulty PCB
Please contact your local retailer or Stocker S.r.l. for assistance and service
For further information about the product, please contact info@stockergarden.com.
Solution
1. Remove the filter from the tank
and rinse it thoroughly
2./3. Send it for repair to an
authorized local retailer or to
Stocker Srl
1./2. Send it for repair to an
authorized local retailer or to
Stocker Srl
3. Check the error message on the
display and check that the nozzles
are fitted correctly.
4. Fill the tank
5. Check programming of the
intervals
1. Charge the battery
2. Replace the battery
3./4. Send it for repair to an
authorized local retailer or to
Stocker Srl
1./4. Send it for repair to an
authorized local retailer or to
Stocker Srl
2. Charge the battery
3. Replace the battery
1. Send it for repair to an
authorized local retailer or to
Stocker Srl
67

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

410411

Inhaltsverzeichnis