6 SIMPLE CALIBRATION

Press [Smpl] and [Print] together whilst the display is counting down from
POWER UP, the display will show "unload CAL." 。
Make sure the top pan is empty and when the stable symbol shows press
[Tare/Zero] to enter calibration mode. The [Unit] key allows you to select a
calibration weight value. See the table as below.
Press [Tare/Zero] to select the correct value
The display will show "LOAD WEI". Place the weight onto the pan and press
[Tare/Zero] when the stable symbol shows. If calibration is correct the display
will show "PASS" and then go back to weighing mode.
If the weight placed on is more than 110% of the calibration weight value selected,
the display will show "FAlL H". If the weight placed on is less than 90% of
calibration weight value selected, the display will show "FAlL L".
Users should recalibrate the balance following the correct operation.+