4.1.3 Percent Weighing
This is used to measure the percentage between the sample weight and a
standard weight.
E.g. Put your standard weight onto the pan and press the [%] key, the display will
show 100.0%.per. This would then be taken as the standard weight. Remove the
standard weight and the reading goes to zero, the display will show 0.0%.
You now place your sample weight onto the pan, the display will show the
percentage of this sample weight in relation to the standard 100% weight.
Press the [%] key again and the display will return to normal weighing mode.
Note: if the standard weight is too small the percentage result may be not
accurate.
4.1.4 Counting mode
• P ut a sample onto the pan and when the reading is stable, press the
[Smpl] key, the display will show "SP 10 pcs";
• P ress the [Unit] key to select sample values from 10, 20, 50, 100 and
200.
• P ress the [Smpl] key again, the display will show the unit weight for 1
piece for 1 second and then show the quantity pcs.
• T o go back to weighing mode, press the [Smpl] key.
4.1.5 Weighing accumulation
A ccumulation of each weighing value can be done, two modes are
available: manual accumulation and automatic accumulation.
4.1.6 Manual accumulation
• I n manual accumulation mode, place a weight onto the pan, after the
stable symbol shows press the [Print] key, the weight will then be
accumulated. The display will show "1 ACC " for 1 second and then
return to weighing mode,
• R emove the weight and the scale will show zero, then place another
weight onto the pan and follow the same procedure to accumulate
7