Herunterladen Inhalt Inhalt Diese Seite drucken

Centrifuge Fc5816 - OHAUS Frontier FC5714 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Frontier FC5714:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-26

9.1.4 Centrifuge FC5816

Model
Speed Range
Maximum RCF
Maximum Capacity(Rotor)
Temperature range(N/A)
Running Time
Noise level (depending on the rotor)
Allowable density at maximum speed
Allowable kinetic energy
Mains power connection AC
Voltage fluctuation
Current consumption
Power consumption
Dimensions (W × D × H)
Net Weight (without rotor)
Shipping Dimensions (W × D × H)
Shipping Weight (without rotor)
Ambient conditions (EN/IEC 61010-1)
Environment
Altitude
Ambient temperature
Max. relative humidity
Overvoltage category (IEC 60364-4-443)
Degree of contamination
Class of protection
Not suitable for use in hazardous environments.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R
200 rpm -15000 rpm;10 rpm/set
21379 x g;10 x g/set
10 sec to 99 hr 59 min 59 sec or continuous
≤ 61 ± 2 dB(A)
230 V ~ 50/60 Hz
± 10 %
2.4 A
530 W
446 x 538 x 354 mm
17.6 x 21.2 x 13.9 in
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
for indoor use only
Use up to an altitude of 2000 m
2°C up to 35 °C
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
FC5816
6 x 250ml
Air cool
1.2 g/ml
34363 Nm
120 V ~ 50/60 Hz
± 10 %
4.2 A
520 W
52 kg
115 lb
77 kg
170 lb
II
2
I

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis